My physiology lab records ECG, pulse and heart sounds with Biopac hardware and its Biopac Student Lab (BSL) software. At the end of the session you get your recording as a file with no extension, which only BSL can open, and BSL is licensed to the lab computers. The lab report, meanwhile, is a stack of tables you fill in by hand: heart rate across three cardiac cycles per condition, P-wave duration, QRS amplitude, R-wave to first heart sound. The analysis isn't hard. It just can only be done on campus.

So I built openbsl: a small Python package that reads the files, and a static browser page that reproduces the part of BSL a lab report needs. There's a live demo running on synthetic data.

The file format

file says the recordings are just data, but the first few hundred bytes give it away: a binary header of big-endian doubles, then per-channel headers with names like ECG and Heart Rate. It's an AcqKnowledge file, the same container Biopac uses in its research software, and bioread already parses it. That part of the problem was already solved:

datafile = bioread.read_file("Keerthik-L05")
[(ch.name, ch.units, ch.samples_per_second) for ch in datafile.channels]
# [('ECG', 'mV', 1000.0), ('Heart Rate', 'BPM', 1000.0)]

The interesting structure is in the event markers, which are the only record of what the subject was doing:

Two details matter downstream. First, the splice: the first R-R interval of every segment spans a pause that isn't in the file. Averaged naively, it produces a phantom 115 BPM beat at the start of a resting segment. The per-segment summaries drop the first interval of each segment for that reason. Second, BSL's heart-rate channel (CH40) is a step function that updates at the end of each R-R interval. The value you read at any point is the rate of the previous beat, which is why the lab manual tells you to "go past the first two cardiac cycles".

Segment labels also carry the hardware serial: Supine, MP36E…. My first cleaner split on the comma, which works for Lesson 5 and silently breaks Lesson 7, whose segments are called things like Seated, left hand in water. It now strips only a trailing , MP… suffix, and there's a test with a comma-containing label.

Trying to automate it first

The lesson 5 report wants P, QRS and T durations and amplitudes, plus PR, QT, ST and TP intervals, for three cardiac cycles. The obvious move is automatic delineation, so I ran NeuroKit2's wavelet delineators on three real recordings from our lab group and checked beat-to-beat consistency within the resting segment (median ± SD, ms):

lead method P duration QRS T QT
clean DWT 68 ± 31 103 ± 2 133 ± 8 302 ± 3
clean CWT 107 ± 9 95 ± 5 163 ± 7 311 ± 5
noisy DWT 72 ± 22 116 ± 20 42 ± 18 273 ± 47
noisy CWT 81 ± 19 110 ± 9 108 ± 34 345 ± 43

On the clean lead the QRS is solid, but the two methods disagree by 40 ms on P duration, and DWT's P-wave jitter is half the P wave. On the noisy lead it falls apart: a 42 ms T wave isn't a T wave. After exercise, with baseline wander and motion artifact, DWT failed to bound the T wave on 51 of 68 beats. Overlaying the raw beats made the reason obvious: the noisy lead's P wave sits at about the amplitude of the noise, and each person's T-wave morphology is different.

The lab's questions are things like "did the QT interval shorten after exercise?" The effect is tens of milliseconds, the same size as the delineator's error bars. An automated number I'd have to check by eye anyway is worse than just measuring by eye, so the tool puts a human on the boundaries, as BSL does.

The one thing that is automated is R-peak detection in the CSV export, where it's reliable: a second-order 5–30 Hz Butterworth band-pass (zero-phase, via filtfilt) to isolate QRS energy, polarity chosen by whichever side has the larger 99th percentile (so inverted leads work), then find_peaks with a threshold at 35% of that percentile and a 300 ms refractory period. On the recordings I checked, its beat-to-beat rate lines up with BSL's own heart-rate channel beat for beat.

The viewer

The viewer is a single static page: an HTML file, two scripts and a generated data.js. openbsl view FILE... writes that folder and opens it. There's no server and nothing leaves your machine, which matters for health data. It loads with classic <script> tags rather than ES modules because browsers block module imports over file://.

Interaction copies BSL, since the lab manual is written against BSL: drag to make an I-beam selection, and the measurement boxes show Delta T, BPM (60 / Delta T), and per-channel P-P (max − min over the selection) and Value. Value is read at the end you dragged to, which is how BSL does it and what the manual's heart-rate procedure depends on. You pick a condition, component and cycle number, and press Enter. The condition is guessed from the nearest marker and the cycle number auto-increments.

Rendering is one canvas per page with a pane per channel. When zoomed out there are many samples per pixel, so each pixel column draws a vertical line from its min to its max. That keeps two minutes at 1 kHz (120,000 points per channel) smooth to pan, and it never hides a QRS spike, as naive downsampling can. The heart-rate channel is stored as change points (index and value) instead of 120,000 samples: about 200 numbers per recording.

Each lesson is one config object in lessons.js:

{
  conditions: ["At rest", "Inhalation", "Exhalation", "After exercise"],
  match: [[/exhal/i, "Exhalation"], [/inhal/i, "Inhalation"], ...],
  components: [{ name: "1st heart sound", ch: /steth/i, tip: "Select across the whole 1st sound" }, ...],
  report(api) { /* annotations -> report tables */ },
}

match maps marker text to conditions and is checked in order. In Lesson 7 "Seated, right hand above head" also contains "seated", so the specific patterns go first. ch says which channel's P-P a component uses (QRS amplitude from the ECG, pulse amplitude from the plethysmograph). report turns annotations into the lab's tables: means over the labelled cycles, pulse-wave speed from a distance you type in, and an increased/decreased/no-change column (±5%) for the heart-sounds lesson. Lessons 5, 6, 7, 16 and 17 are configured. Anything else falls back to one condition per segment.

Testing without publishing anyone's ECG

I can't put real recordings in a public repo or demo, so openbsl.synthetic generates them. Each beat is a sum of five Gaussians (P, Q, R, S, T) in the style of McSharry et al.'s dynamical ECG model. The T wave's position and width scale with √RR, so QT shortens as heart rate rises. The rhythm has respiratory sinus arrhythmia during deep breathing (HR rises on inspiration, with inhale/exhale markers placed to match) and an exponential recovery after exercise. There's also noise and baseline wander. Lesson 7 gets a pulse wave with a dicrotic bump, and Lesson 17 gets windowed 45 and 60 Hz bursts for S1 and S2.

Because the true beat times are known, the detector is tested against ground truth instead of against itself: every beat found within 5 ms at 48, 75 and 140 BPM, on an inverted lead too, and the demo's 64 and 72 BPM segments recovered to within 0.5 BPM. The report tables are tested in Node against hand-computed values. For example, 40 + 60 cm over a 250 ms R-to-pulse delay has to come out as 400.0 cm/s, and a 5% rise exactly on the boundary has to read "no change".

Update: Lessons 6 and 16, where automation does work

October 2026. Two more lessons came along, and both turned out to be the exception to the section above: the measurement they need is large and sharp enough to automate, as long as you're careful about which feature you trust.

Lesson 6: the mean electrical axis

Lesson 6 records Lead I and Lead III, has BSL compute Lead II, and asks you to confirm Einthoven's law (I + III = II), then to estimate the heart's mean electrical axis by drawing the Lead I and Lead III QRS amplitudes on a protractor diagram. The measurement is BSL's Delta: the last point of the selection minus the first, dragged from the isoelectric line (the midpoint between the previous T wave and this P wave) to the peak, keeping the sign. The viewer didn't have Delta, so it got one.

The protractor construction (perpendiculars from the two lead vectors, meeting at the tip of the resultant) has a closed form. With Lead I along 0° and Lead III along +120°:

x = Lead I,   y = (2 · Lead III + Lead I) / √3,   axis = atan2(y, x)

openbsl axis measures Q, R and S on each lead for one representative beat per condition and writes the report's three graphs as SVG vectorgrams, with the perpendiculars drawn in so you can see the construction. The tests build synthetic leads by projecting known heart vectors onto each lead, so the recovered axis is checked against the angle it was built from.

Two things went wrong on the way, both instructive:

Lesson 16: finding Korotkoff sounds

Lesson 16 is blood pressure: a cuff with a pressure transducer, a stethoscope with a microphone in its head, and an ECG. One person listens and presses a key at the first sound (systolic) and when the sounds stop (diastolic). The report then asks you to find the same two points on the microphone trace and compare.

On the microphone trace the sounds don't stand out. The stethoscope sits under the cuff and hears every rustle, and plenty of bumps are as loud as the sounds. What distinguishes a Korotkoff sound is timing: the artery snaps open once per beat, so each sound arrives at nearly the same delay after the R wave, and that delay drifts smoothly (from about 400 ms near systolic, when the artery only opens at the very peak of the pulse, to about 270 ms near diastolic). So the detector scores each beat by how far the 20–150 Hz envelope rises above that beat's own pre-R noise, 120–480 ms after R. It then keeps runs of at least three beats whose delays agree to within 60 ms. Random bumps almost never line up like that.

It reports a result only if the sounds appear. There have to be quiet beats before the first one, the sounds have to be at least 1.5× louder than those beats, and they have to fill most of the beats between first and last. Each of those gates exists because of a real recording:

On synthetic recordings with known pressures the detector is within 3.5 mmHg in 57 of 60 trials, and makes no detections on noise-only recordings.

On real data, the microphone heard systolic 6–28 mmHg before the person listening, in every trial. Diastolic agreed within a few mmHg. In the lying-down trials the first sounds are faint taps that grow over several beats before anyone presses the key. That's phase I of the Korotkoff sounds, and it's exactly what the report's first question is fishing for.

My own recording had no sounds at all on the microphone. To check whether they were just faint, I averaged all 167 beats between the markers against 235 beats above systolic. There was no difference in any band. Then I took the other subject's real sounds, scaled them down, and added them to my noise: at 10% of their loudness the same test still found them. So whatever reached the microphone in my trials was at least ten times quieter.

Then I looked at the raw traces, which I should have done first. The cuff pressure isn't a smooth ramp: each heartbeat adds a small ripple as the artery pushes on the cuff. In a good recording that ripple grows as the cuff deflates and peaks between systolic and diastolic, which is what automatic blood pressure monitors measure. In mine it was largest at the top of the cuff range and shrank all the way down. That's the pulse hammering on the upper edge of a cuff that isn't compressing the artery under it. A misplaced cuff and stethoscope explain the missing sounds; quiet arteries don't. The fix is the procedure's own placement step, not better signal processing.

What's missing

Code is on GitHub, MIT licensed.